Abstract
Context Hypertension Canada recommends home blood pressure monitors (HBPM) for diagnosing and managing hypertension. However, 29% of HBPMs are inaccurate by 10 mmHg or more. This makes determining HBPM accuracy of great clinical importance, but there is no established method of doing so in the office. Serial blood pressure (BP) measurements can be attempted, but a large number is needed to rule out transient changes in BP.
Objective To introduce the novel approach of using a T-Connector between a patient’s HBPM and their physician’s office sphygmomanometer to permit a simultaneous blood pressure reading.
Study Design 89 participants had the accuracy of their personal HBPMs assessed by 1) a family physician using our T-Connector Approach and 2) the gold standard method called Dual Observer Auscultation, which involves 2 physicians alternating between the HBPM and a mercury manometer for 9 readings.
Setting Kaye Edmonton Clinic Family Medicine Clinic and a hypertension research lab.
Population Adults who own a HBPM.
Instrument The T-Connector Approach uses latex tubing, a plastic barbed tee piece, and a metal connector from the primary care clinic (the piece used to switch blood pressure cuffs) to connect a patient’s HBPM to their physician’s office sphygmomanometer.
Outcome Measures an inaccurate HBPM was defined as providing a blood pressure reading ≥ 10 mmHg different than the T-Connector or dual observer auscultation methods. We primarily compared the ability of the T-Connector to detect inaccurate monitors compared to the dual observer auscultation (DOA) gold standard method.
Results Compared to DOA, the T-Connector’s assessment of HBPM error was an average of 2.97 mmHg lower, 95% limits of agreement were -10.72 to 16.65 mmHg, and the intraclass correlation coefficient (ICC) was 0.47. Sensitivity and specificity for detecting an inaccurate monitor were 0.33 and 0.97 respectively, with a positive predictive value (PPV) of 0.75, and a negative predictive value (NPV) of 0.85. Cohen’s kappa was 0.39.
Conclusion Based on the Cohen’s kappa of 0.39 (fair agreement), PPV of 0.75, and NPV of 0.85 there appears to be reasonable utility for using a T-Connector to qualitatively declare a HBPM to be inaccurate for its user. However, the sensitivity was low (0.33) and the ICC being <0.5 indicates poor agreement for quantitative assessment of the mmHg of error.
- © 2024 Annals of Family Medicine, Inc. For the private, noncommercial use of one individual user of the Web site. All other rights reserved.